Description
Overview
The GE MIC 10 is a compact tool that allows for quick and convenient,on-site hardness testing virtually anywhere.
Unlike traditional optical evaluation with conventional hardness testers, the MIC 10 electronically measures the material’s surface and instantly displays the results. Measurements can be made on various materials, objects of various shapes and sizes, and even heat-treated surfaces. Erroneous single values can also be edited without interrupting the measurement process. Keep your inspection on track with adjustable visual and audible alarms when reaching critical measurement values.
The MIC 10 is automatically calibrated for low and non-alloyed steel materials but can easily be re-calibrated for other materials. The instrument operation can just as easily be customized for application-specific hardness testing requirements, allowing the inspector to reduce key presses by disabling unnecessary functions and simplifying operations. The narrow probes also give the inspector access to difficult-to-reach areas.
In addition to the internal storage included on the MIC 10, instrument settings and calibrations can also be stored on a memory card and automatically recalled when measuring specific materials. Expand your capabilities using the UltraHARD software to make further data processing such as evaluations, statistics, and documentation.
Features
- On-site hardness testing according to the Ultrasonic-Contact-Impedance (UCI)method.
- Standardized according to ASTM A 1038.
- Probes with test loads from 1N (HV0.1) to 98N (HV10).
- Automatic conversion into different hardness scales and tensile strength.
- Display of single value or average from a test series.
- Data Logger version with internal data memory for measurement data and instrument settings.
Specifications
General | ||
Vickers penetration with indentation evaluation by the UCI method under load (diamond angle 136°) |
||
Range | 20 – 1740 HV | |
Conversion | HV, HB, HRC, HRB | |
Display | 4-digit LCD with switchable backlight | |
Statistics | Display of the average value. | |
For version MIC 10 DL: printout with maximum, minimum, average value, absolute and relative range, absolute and relative standard deviation. |
||
Power | 2 x 1.5V AA batteries | |
Operating Time | ~15 hours without backlight | |
Operating Environment | Operation | -15°C to 55°C(5°F to 131°F) | |
Storage | -20°C to 60°C (-4°F to 140°F) | ||
Weight | ~300 g (~ 0.7 lb) | |
Dimensions | 160 mm W x 70 mm H x 45 mm D (6.3″ W x 2.8″ H x 1.8″ D) | |
Interface (for version MIC 10 DL) | RS-232C bidirectional | |
Data Logger (for version MIC 10 DL) | Internal memory for up to 1800 measurments, memory card for up to 590 measurements. Dependent on the number of measurements per set. Warning given with memory overflow. |
|
Statistics | Display of the average value. | |
For version MIC 10 DL: printout with maximum, minimum, average value, absolute and relative range, absolute and relative standard deviation. |
Reviews
There are no reviews yet.